The VELOCICALC® 9565 Multi-Function Ventilation Meter features
a menu-driven user interface for easy operation in your local language.
On-screen prompts and step-by-step instructions guide the user through
instrument setup, operation and field calibration. The 9565 also
features an ergonomic, overmolded case design with probe holder
and a keypad lockout to prevent tampering during unattended use.
The 9565 model includes the Straight Air Velocity Probe 964 and
is designed to work with a wide range of plug-in probes.
The probes can be ordered at any time and include a data sheet
with certificate of traceability. When it is time for servicing,
only the probe needs to be returned since all the calibration data
is stored within the probe.
|
|
Features and Benefits
- Straight Air Velocity Probe 964 measure air velocity,
temperature and relative humidity
- Includes differential pressure sensor
- Best-in-class air velocity accuracy
- Optional smart plug-in probes, including VOC, CO2,
and rotating vane probes
- Accommodates up to two K-alloy thermocouples
- Large graphic display
- Displays up to five measurements simultaneously
- On-screen messages and instructions
Program for local language
- Intuitive menu structure allows for ease of use and setup
- Manual or continuous data logging
- Name test IDs meaningful to you
- Multiple data logging formats
- Bluetooth communications for transferring data or remote polling
- Includes TRAKPRO and LogDat2 downloading software
with USB cable
|
|
Features
|
9565
|
9565-A
|
9565-P
No Probes |
9565-X
No Probes
|
| Probe
that measures velocity, temperature and humidity |
straight
Model 964 |
articulated
Model 966
|
optional |
optional |
| Pressure
measurement |
x
|
x
|
x
|
No
|
| Calculates,
flow, wet bulb, dew point, standard, actual |
x
|
x
|
x
|
x
|
| Optional
temperature and velocity probe |
x
|
x
|
x
|
x
|
| Optional
rotating vane velocity probe |
x
|
x
|
x
|
x
|
| Optional
IAQ probes (CO, temperature, humidity, CO2) |
x
|
x
|
x
|
x
|
| Optional
VOC probes (VOC Concentration, temperature) |
x
|
x
|
x
|
x
|
| Data
logging (manual, auto save, continuous) |
x
|
x
|
x
|
x
|
| Data
logging software |
x
|
x
|
x
|
x
|
| Optional
Blue Tooth printer |
x
|
x
|
x
|
x
|
Specifications
VelociICalc Models 9565, 9565-A, 9565-P, 9565-X and Optional Probes
| Velocity
(TA Probe Models 960, 962, 964, 966) |
| Range
|
0
to 9,999 ft/min (0 to 50 m/s) |
| Accuracy1,2
|
±3%
of reading or ±3 ft/min (±0.015 m/s), whichever
is greater |
| Resolution
|
1
ft/min (0.01 m/s) |
| Velocity
(Pitot Tube for Meter Models 9565, 9565-A, 9565-P) |
| Range3
|
250
to 15,500 ft/min (1.27 to 78.7 m/s) |
| Accuracy4
|
±1.5%
at 2,000 ft/min (10.16 m/s) |
| Resolution
|
1
ft/min (0.01 m/s) |
| Velocity
(Rotating Vane Probe Model 995) |
| Range |
50
to 6,000 ft/min (0.25 to 30 m/s) |
| Accuracy |
±1% of reading ±4 ft/min (±0.02 m/s |
| Resolution
|
1
ft/min (0.01 m/s) |
| Duct
Size |
| Dimensions
|
1
to 250 inches in increments of 0.1 in.
(1 to 635 cm in increments of 0.1 cm) |
| Volumetric
Flow Rate |
| Range
|
Actual
range is a function of velocity, pressure, duct size, and K
factor |
| Temperature
(TA Probe Models 964 and 966, IAQ Probe Models 980 and 982) |
| Range
|
14
to 140°F (-10 to 60°C) |
| Accuracy5 |
±0.5°F
(±0.3°C) |
| Resolution
|
0.1°F
(0.1°C) |
| Temperature
(TA Probe Models 960 and 962) |
| Range
|
0
to 200°F (-18 to 93°C) |
| Accuracy5
|
±0.5°F
(±0.3°C) |
| Resolution
|
0.1°F
(0.1°C) |
| Temperature
(Rotating Vane Probe Model 995) |
| Range
|
32
to 140°F (0 to 60°C) |
| Accuracy
|
±2.0°F
(±1.0°C) |
| Resolution
|
0.1°F
(0.1°C) |
Temperature
(K-Alloy Thermocouple Surface Probe Model 792) |
| Range
|
-40
to 1200°F (-40 to 650°C) |
| Accuracy
|
±0.1%
of reading +2°F (±0.056% of reading +1.1°C) |
| Resolution
|
0.1°F
(0.1°C) |
| Temperature
(K-Alloy Thermocouple Air Probe Model 794) |
| Range
|
-40
to 1600°F (-40 to 870°C) |
| Accuracy
|
±0.1%
of reading +2°F
(±0.056% of reading +1.1°C) |
| Resolution
|
0.1°F
(0.1°C) |
| Relative
Humidity (TA Probe Models 964 and 966, IAQ Probe Models 980
and 982) |
| Range
|
0
to 95% RH |
| Accuracy6
|
±3%
RH |
| Resolution |
0.1%
RH |
| |
| Optional
Probes for VELOCICALC 9565 Series Multi-Functional Anemometers
|
| 960
|
Air
Velocity and Temperature, straight probe |
| 962
|
Air
Velocity and Temperature, articulating probe |
| 964
|
Air
Velocity, Temperature, and Humidity, straight probe |
| 966
|
Air
Velocity, Temperature, and Humidity, articulating probe |
| 995
|
4
in. (100 mm) Rotating Vane probe |
| 792
|
Surface
Temperature probe |
| 794
|
Air
Temperature probe |
| 980
|
Indoor
Air Quality probe |
| 982
|
Indoor
Air Quality probe, with CO |
| 984
|
Low
Concentration VOC probe, with temperature |
| 985
|
High
Concentration VOC probe, with temperature |
| 986
|
Low
Concentration VOC probe, with temp, CO2, Rh |
| 986 |
High
Concentration VOC probe, with temp, CO2, Rh |
|
| Carbon
Monoxide (IAQ Probe with CO Model
982)
|
| Range
|
0
to 500 ppm |
| Accuracy7
|
±3%
of reading or ±3 ppm, whichever is greater |
| Resolution
|
0.1
ppm |
| Carbon
Dioxide (IAQ Probe Models 980 and 982) |
| Range
|
0
to 5000 ppm |
| Accuracy8
|
±3%
of reading or ±50 ppm, whichever is greater |
| Resolution
|
1
ppm |
Static/Differential
Pressure
(Meter Models 9565, 9565-A, 9565-P) |
| Range9
|
-15
to +15 in. H2O (-28.0 to +28.0 mm Hg,-3735 to +3735 Pa) |
| Accuracy
|
±1%
of reading ±0.005 in. H2O (±0.01 mm Hg, ±1
Pa) |
| Resolution
|
0.001
in. H2O (0.1 Pa, 0.01 mm Hg) |
| Barometric
Pressure |
| Range
|
20.36
to 36.648 in. Hg
(517.15 to 930.87 mm Hg) |
| Accuracy
|
±2%
of reading |
| Instrument
Temperature Range |
| Operating
(Electronics) |
40
to 113°F (5 to 45°C) |
| Storage
|
-4
to 140°F (-20 to 60°C) |
| Data
Storage Capabilities |
| Range
|
26,500+
samples and 100 test IDs |
| Logging
Interval |
1
second to 1 hour |
| Time
Constant |
User
selectable |
| Meter
Dimensions |
| External
Meter Dimensions |
3.8
in. x 8.3 in. x 2.1 in. (9.7 cm x 21.1 cm x 5.3 cm) |
| Meter
Weight with Batteries |
0.8
lbs. (0.36 kg) |
| 960
& 962 Meter Probe Dimensions |
| Probe
length |
40
in. (101.6 cm) |
| Probe
Diameter of Tip |
0.28
in. (7.0 mm) |
| Probe
Diameter of Base |
0.51
in. (13.0 mm) |
| 962
Articulating Probe Dimensions |
| Articulating
Section Length |
6
in. (15.2 cm) |
| Diameter
of Articulating Knuckle |
0.38
in. (9.5 mm |
| |
| Power
Requirements |
Four
AA-size batteries or AC adapter |
Optional
Probes
(Click
to enlarge photo)
|
Probe
Suitability Matrix
|
|
9565
|
9565-A
|
9565-P
|
9565-X
|
| Probe
that measures velocity, temperature, and humidity |
includes
964 probe |
includes
966 probe |
optional
|
optional
|
| Pressure
measurement |
x
|
x
|
x
|
No
|
| Calculates
flow, wet bulb, dew point, standard/actual |
x
|
x
|
optional
|
optional
|
| Optional
velocity and temperature probe |
x
|
x
|
x
|
x
|
| Optional
rotating vane probe |
x
|
x
|
x
|
x
|
| Optional
IAQ probes (CO2, temperature, humidity, CO) |
x
|
x
|
x
|
x
|
| Data
Logging (manual, auto save continuous) |
x
|
x
|
x
|
x
|
| Data
logging software |
x
|
x
|
x
|
x
|
| Optional
Bluetooth printer |
x
|
x
|
x
|
x
|
| Optional
Bluetooth printer |
x
|
x
|
x
|
x
|
Footnotes
- Temperature
compensated over an air temperature range of 40 to 150°F
(5 to 65°C).
- The
accuracy statement begins at 30 ft/min through 9,999 ft/min
(0.15 m/s through 50 m/s).
- Pressure
velocity measurements are not recommended below 1000 ft/min
(5 m/s) and are best
suited to velocities over 2,000 ft/min (10.00 m/s). Range can
vary depending on barometric
pressure.
- Accuracy
is a function of converting pressure to velocity. Conversion
accuracy improves when actual
pressure values increase.
- Accuracy
with instrument case at 77°F (25°C), add uncertainty
of 0.05°F/°F (0.03°C/°C) for change
in instrument temperature.
- Accuracy
with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F
(0.2% RH/°C) for change in
probe temperature. Includes 1% hysteresis.
- At
77°F (25°C). Add uncertainty of ±0.2%/°F
(0.36%/°C) for change in temperature.
- At
calibration temperature. Add uncertainty of ±0.28%/°F
(0.5%/°C) for change in temperature.
- Overpressure
range = 190 in. H2O (360 mmHg, 48 kPa).
Specifications
subject to change without notice.
|
|
Return
to Brandt Instruments
For
more information Email
Brandt
Instruments, Inc.
18568 Oak Grove Parkway
Prairieville, LA 70769
1-800-337-6291
Phone: 225-673-6776
Fax: 225-673-6725
Copyright © 2002-2012 Brandt Instruments,
Inc.
Notice:
This publication is intended for general information only and
should not be used as a substitute for reviewing applicable
government regulations, equipment operating instructions, or
legal standards. The information contained in this document
should not be construed as legal advice or opinion nor as a
final authority on legal or regulatory procedures. Brandt Instruments,
Inc. reserves the right to make any changes to information,
features, and specifications of any product without notice.