Biosystems PhD Ultra Confined Space
Meter
Sperian Instruments now Honeywell Analytics
 |
DATALOGGING
GAS DETECTOR
The PhD Ultra portable gas
detector monitors up to four atmospheric hazards simultaneously:
oxygen, combustible gas, and up to two additional toxic
gases.
|
"Smart Sensors" available
include:
Oxygen (O2)
LEL - combustible gases
Carbon Monoxide
(CO)
Hydrogen Sulfide
(H2S)
Sulfur Dioxide
(SO2)
Dual purpose "CO Plus"
Sensor for
simultaneously measuring
CO & H2S |
Hydrogen Cyanide (HCN)
Nitric Oxide
(NO)
Nitrogen Dioxide
(NO2)
Ammonia
(NH3)
Chlorine
(Cl2) |
The PhD Ultra automatically
logs gas readings, temperature, location, user
ID, and other key information at intervals set by the user from
one second to one hour. A real-time clock allows the instrument
to recordturn-on/turnoff times, calibration dates for each sensor,
and even display
custom "service due" messages! Automatically download to
a serial printer for direct generation of reports, or interface
with a computer with the datalink software for more advanced report
formats.All changes in setup parameters can be done easily through
the instrument keypad or computer interface.
PhD Ultra options include
your choice of two snap-in battery packs. The NiCad
pack option comes complete with a fast charger which allows recharging
in less than two hours. (Available in 110v, 220v, or 12v)The
alkaline pack holds a set of three AA disposable batteries. Either
pack provides up to 12 hours of continuous operation.
The optional slip-on "Smart"
sample-draw pump is powered and electronically
monitored directly by the instrument. PhD Ultra detectors are
also available in the form of complete kits which include all
standard accessories as
well as calibration gas, regulator, and waterproof carrying case.
True
"One Button" operation: All procedures necessary for day-to-day
operation, including automatic calibration adjustment, are controlled
through the single On/Off "Mode" button.
Smart
sensors: Biosystems' revolutionary new "Smart Sensors" are
individually equipped with their own EEPROM memory chip. Besides
storing information such as type, serial number and alarm settings,
the chip also stores calibration adjustments and sensor-specific
temperature compensation curves. Every time the sensor is calibrated
it gets smarter, and accuracy is continuously improved.
Automatic
record keeping: The PhD Ultra logs data automatically, making
record keeping a snap. The instrument remembers (logs) gas readings,
temperature, location, user ID, and other important information.
Interchangeable
NiCad and disposable alkaline battery packs: PhD Ultra batteries
are housed in two types of snap-in battery packs. Either pack
provides up to 12 hours of continuous operation. The NiCad pack
can be recharged while installed in the PhD Ultra, or may be removed
for separate recharging in the PhD Ultra "Datadock" fast charger.
Automatic
calibration adjustment: "One Button Auto Cal" is available
any time while in normal operation. All you have to do is press
the "Mode" button, attach the adaptor, and flow gas to the sensors.
All the adjustments are made automatically.
Small
package; rugged design: The PhD Ultra is completely sealed
against the elements. The rugged steel-filled polycarbonate case
is designed to minimize the effects of radio frequency interference.
Automatic
setup: Instrument setup is automatic; just plug in the new
sensor. The PhD Ultra reviews all the information stored on the
sensor chip, notes the changes, displays the correct readings,
and assigns the proper alarms.
Multi-sensor
capability: One to four sensors: O2, LEL, and up
to two toxic sensors: CO, H2S, S02,Cl2,
NH3, HCN, NO2 and NO, or Biosystems' dual
purpose "CO Plus" sensor for the simultaneous measurement of both
CO and H2S.
Sample
draw options: Every unit comes complete with a manual sample
draw kit. The optional continuous sample pump draws power directly
from the PhD Ultra and operates whenever the instrument is turned
on. The PhD Ultra directly monitors the pump for proper operation.
|
|
Specifications
& Programmable Features Configurations
Calibration
Materials Accessories Sensors
The PhD
Ultra is designed
to be
the heart and soul of
your record
keeping program
- Record
keeping is automatic: The PhD Ultra automatically
remembers (logs) gas readings, turn-on/turn-off times, temperature,
battery condition, calibration dates, types of sensors currently
installed, sensor serial numbers, warranty expiration, service
due dates, current alarm settings and more.
- Datalogging
"transparent": It is not necessary to do anything
special to begin logging data. Simply turning on the instrument
causes it to begin recording. Although it is not necessary to
make use of the information being stored, it is there in the
event it is needed.
- Gas
Detector and Database Software: Biosystems' "Datalink"
Gas Detector and Database Software allows for two-way communication
between your PhD Ultra and an MS DOSTM personal computer.
- One
button is all it takes: Biosystems' Datalink software
has been designed to make programming, downloading, and data
analysis as easy as pushing a button. The software also allows
you to customize your instrument by "filling out" an electronic
form right on your computer screen.
|
- Name
and location ID: You can even add the name of the
PhD Ultra user and the location where the monitoring occurred
by using built-in push-buttons to scroll through a list and
make a selection.
- Analyze
and organize your monitoring records: Biosystems'
Datalink software allows both direct examination of recorded
data as well as easy export to other data management programs.
Besides examining data from specific sessions in the form of
quick summaries, graphs or tables, previously down-loaded records
may be searched by date, exceedence values, location, user name,
sensor type, as well as many other "sorting" parameters.
- Review
your data right on the spot: You don't need a computer
to review your data. Session data can be displayed directly
on the instrument LCD, or downloaded straight to a serial printer.
- Smart
sensor data: Review information directly from the
sensor EEPROMs to verify sensor serial numbers, warranty expiration
dates, and the 8 most recent zero and span calibration dates;
all on a sensor-specific basis!
- Sophistication
without complication: For all the sophistication
of the PhD Ultra, the bottom line is still ease of use. Making
use of these powerful software and database tools is as simple
as using your computer "mouse" to pick your choice from the
pull-down menu of options.
|
Powerful Tools That Are
Easy To Use
Comprehensive
reports with the push of a button:
Recorded information may
be used for a variety of purposes. Data may be displayed and reviewed
in detail through the computer monitor screen, or used to automatically
generate and print reports, tables and graphs of time history exposure
data. It is also possible to export records to other software applications
in the form of ASCII, Lotus compatible, or "comma delimited" database
files. Another option is to simply retain downloaded records of
your gas detection monitoring program for future use. |
|
Automatic
Downloading:
With the PhD Ultra downloading
stored data is a snap! Simply place the PhD Ultra in the slip-in
Datadock cradle, connect the cradle to your computer or serial printer,
and press enter.
- Summary
information:
- User ID
- Monitoring location
- Start / stop times
- Session duration
- Peak high and low
temperatures
- Datalogging interval
- Specific
session data:
- Sensor type and
serial numbers
- Peak high and low
readings
- STEL, TWA and average
- Calibration dates
- Complete
time history data
- Print
graphs automatically
|
| Sample
Drawing:
Every PhD Ultra
comes complete with a manual (squeeze bulb) sample draw kit with
10 feet of sample tubing. The sample draw adaptor also serves
as the calibration adaptor used to introduce test gas to the sensors.
Continuous
Sample Pump:
An optional slip-on
UL®, CSA Classified sample draw pump is available
for situations requiring continuous hands-free remote sampling.
The sample pump draws power directly from the PhD Ultra battery.
The pump automatically operates whenever the instrument is turned
on. The PhD Ultra microprocessor directly monitors the pump for
proper operation. The pump contains a unique pressure sensor which
detects water or other fluids being sucked into the hose, and
immediately shuts off the pump in order to prevent damage.
Compact
size; rugged design:
The PhD Ultra is housed
in a super-rugged steel-filled polycarbonate case designed to
stand up to the most demanding environments. The completely sealed
water-resistant case is protected against dust and liquid intrusion
and has been designed to minimize the effects of radio frequency
interference. Individual LED alarm lights for each gas being measured
and an automatic backlight that turns on in dim light means that
the PhD Ultra is always convenient to use; even in the dark!
|
|
Sophistication
Without Complication
Front
Panel is Easy to See and Read:
Three operating modes
make the PhD Ultra easy to use ...
Text Only
Mode:
This
mode provides a quick indication that all systems are GO until an
alarm is activated. The Text Only mode makes the PhD Ultra the easiest
to use instrument on the market, keeping training requirements to
a minimum.
Basic
Mode:
In
this mode readings are given in numbers. "One Button Auto Calibration"
can be used anytime the instrument is operated in the Basic Mode
to verify that the instrument is accurate.
Technician
Mode:
This
is the mode for advanced users. In this mode you have access to
all advanced functions and displays as well as "One Button Auto
Calibration" adjustment.
In normal operation the
display identifies and shows readings for all the gases being
measured at the same time. Simply push the Mode button to view
Peak Exposures, Short Term Exposure Levels, and Time Weighted
Averages. In the Technician Mode you can change alarm settings
as well as make use of optional setup choices.
|
Smart
Sensors Give The PhD Ultra a Touch of Genius:
Each Biosystems "Smart Sensor"
is individually equipped with its own memory chip. Besides storing
information such as type, serial number and alarm settings, the
chip also stores calibration adjustments and sensor-specific temperature
compensation curves. This information is read and updated every
time the instrument is turned on. Every time the instrument is calibrated
the sensors get smarter, and accuracy is continuously improved.
Choose from substance specific sensors for the detection of CO,
H2S,
SO2, CL2,
NH3, HCN, NO and NO2 or Biosystems' dual purpose
"CO Plus" sensor for the simultaneous measurement of both CO and
H2S.
Alarms:
PhD Ultra alarms are user
adjustable and may be set anywhere within the range of the sensor.
(Factory settings may be restored at any time by using the "alarm"
push-button.) When an alarm point is exceeded a loud audible alarm
sounds, and an individual bright red LED alarm light for each
affected sensor blinks.
Convenient
Kits:
PhD Ultra detectors are
available in the form of complete kits.
Confined Space Kits
include a PhD Ultra detector with all standard accessories as
well as calibration gas, regulator and waterproof carrying case.
|
|
Calibration
is Easy and Automatic:
Biosystems' "One Button"
logic makes calibration a snap. All you have to do is press the
"Mode" button, attach the adaptor, and flow gas to the sensors.
All the adjustments are made automatically.
"One Button Auto Calibration"
is available any time while in normal operation.
You start by using the
Mode button to make a fresh air "Auto Zero" adjustment. When the
zero adjustment has been completed the PhD Ultra will prompt you
to do a "Span Calibration."
Simply press the Mode
button, and flow gas to the sensors.
The PhD Ultra will wait
for gas to be applied, show you the current readings, and then
(if necessary) automatically adjust the sensors.
The screen will show the
reading for each sensor as it is adjusted, then let you know when
calibration is complete.
Datalink Kits include
Biosystems Datalink Gas Detector and Database Software, computer
interface cable, and owners manual. Kits may be ordered either
with or without the "Datadock" fast charger.
|
Battery
Options:
PhD Ultra batteries are housed in two types of interchangeable snap-in
battery packs. Either pack provides up to 12 hours of continuous
operation. The alkaline pack holds a set of 3 AA disposable batteries.
The NiCad pack can be recharged while installed in the PhD Ultra
or may be removed from the instrument for separate recharging in
the PhD Ultra "Datadock" fast charger. The intrinsically safe battery
packs are designed so that they may be removed or replaced even
while the instrument is being operated in a hazardous location. |
PhD
Ultra "Datadock" Fast Charger:
When ordered with a NiCad battery pack, the standard accessories
include a PhD Ultra "Datadock" fast charger which allows charging
of the battery either in or out of the instrument in less than two
hours. Built-in LED lights indicate charging status. When fast-charging
is complete the charger automatically converts to a trickle charge
in order to avoid damage to the battery pack due to overcharging. |
Training Video:
A FREE training video is included with
every PhD Ultra shipped. |
Accessories:Every
PhD Ultra is shipped complete with leather case, carrying strap, hand-aspirated
sample draw kit with 10' of tubing, quick reference card & owner's
manual. The sample draw kit includes sample draw / calibration adaptor,
squeeze bulb, tubing, sample probe and spare sample probe filters.
"Alkaline" PhD
Ultra detectors include an alkaline battery pack and set of 3
AA disposable batteries.
"NiCad" PhD Ultra
detectors include a NiCad battery pack and slip-in PhD Ultra "Datadock"
combination fast charger/computer interface cradle, as well as
spare alkaline battery pack and set of batteries.
"Pro" PhD Ultra
detectors include a NiCad battery pack, slip-in PhD Ultra "Datadock"
fast charger/computer interface cradle, "Datalink" gas detector
and database software, computer interface cable, and spare alkaline
battery pack and set of batteries.
|
Specifications
& Programmable Features
| Power: |
Snap-in alkaline
or rechargeable NiCad battery packs |
| Battery life: |
12 plus hours |
| Weight: |
NiCad: 22 oz. (623g)
Alkaline: 19.5 oz.
(538g) |
| Dimensions: |
6.75 inches (17cm)
x 3.75 inches (9.5cm) x 2 inches (5cm) |
| Intrinsic safety: |
Canadian Standards
Association and U.L. Classified Class 1, Division 1, Groups
A, B, C & D |
| Alarms (audible): |
Gas alarm set-points,
low battery, electronic fault, missing sensor |
| Alarms (visual): |
Bright red LED for
each gas measured |
| Display: |
Supertwist LCD, simultaneously
shows readings and messages for all gases measured |
| Display back light: |
Automatic activation
in dim light |
| Sensors: |
One to four "Smart
Sensors": oxygen, LEL combustible, and up to two toxic gases |
| Smart Sensor features: |
Each sensor stores
type of sensor, serial number, alarm settings, calibration
dates, and other information |
| Calibration: |
Fully automatic "zero"
and "span" calibration adjustment |
| Calibration dates: |
Automatically retains
8 most recent "zero" and "span" adjustment dates independently
for each sensor |
| Automatic setup: |
Instrument automatically
updates and notes changes in sensors installed |
| Setup options: |
Change settings through
instrument keypad or computer interface |
|
| Sample draw: |
Standard accessories
include manual (squeeze bulb) sample draw assembly; motorized
sample draw pump optionally available |
| Datalogging: |
Standard, automatic
when instrument is turned on |
| Real-time clock: |
Automatically time-stamps
datalogged information |
| Operating modes: |
Technician / Basic
/ Text Only |
| Programmable alarm
set-points: |
Oxygen: low/ high;
Combustible: high; Toxic: TWA/STEL/ Ceiling |
| Programmable service
due dates: |
User enabled, independently
assigned for each sensor |
| Latching peak
alarm: |
Enable / disable
by user |
| Display resolution: |
0.1 ppm / 1.0 ppm |
| Security beep: |
Enable / disable
by user |
| Security beep
interval: |
User selectable,
any interval up to 1 hour |
| Datalogging recording
interval: |
1 second to 1 hour |
| Data storage capacity: |
3600 intervals, up
to 45 monitoring sessions |
| User ID list: |
Store up to 10 user
names in instrument memory |
| User location
list: |
Store up to 15 locations
in instrument memory |
| Computer interface: |
Automatic data downloading
with Biosystems' Datalink software |
| Serial printer
interface: |
Automatic data downloading
to serial printer for direct generation of reports |
| Date and time: |
Programmable MM/DD/YY,
HH/MM/SS, user adjustable |
| Elapsed monitoring
timer: |
Records length of
current monitoring session |
|
| PhD Ultra Ordering
Information To
order the PhD Ultra equipped with an "Alkaline" battery pack add
an "A" to the part number. To order as "NiCad" add "N". To order
as a "Pro" version detector add 'P". To order any version instrument
as a complete Confined Space Kit add "C". An optional metal-plated
instrument housing is available for situations requiring additional
protection against RFI. To order the PhD Ultra detector housed in
a metal-plated case add "R" to the part number. All instruments
come complete with leather case, belt clip, shoulder strap, manually-aspirated
sample pump, quick-reference card, owners manual, and comprehensive
training video.
Please Note: Since
the PhD Ultra is designed for almost every combination of oxygen,
LEL and up to two of our eight currently available toxic sensors,
many sensor configurations are possible. The following table shows
only the most common. Please call Biosystems for pricing and ordering
information on other available configurations.
|
Common
PhD Ultra Detector Configurations
| 54-02-30000 |
PhD Ultra with LEL
/ O2 |
| 54-02-30100 |
PhD Ultra with LEL
/ O2 / CO |
| 54-02-30200 |
PhD Ultra with LEL
/ O2 / H2S |
| 54-02-30500 |
PhD Ultra with LEL
/ O2 / "CO Plus" |
| 54-02-30102 |
PhD Ultra with LEL
/ O2 / CO / H2S |
| 54-02-30103 |
PhD Ultra with LEL
/ O2 / CO / SO2 |
| 54-02-30104 |
PhD Ultra with LEL
/ O2 / CO / NH3 (CALL FACTORY) |
| 54-02-30801 |
PhD Ultra with LEL
/ O2 / CO / CL2 |
| 54-02-30901 |
PhD Ultra with LEL
/ O2 / CO / NO2 |
| 54-02-30906 |
PhD Ultra with LEL
/ O2 / NO2 / NO |
|
PhD
Ultra Calibration Materials
| 54-05-K0501 |
Calibration kit includes
foam-lined carrying case with room for up to three cylinders
of calibration gas. Kit contains one 1.0 LPM fixed flow rate
regulator and other fittings. Kit does not include gas. |
| 54-9031 |
Methane 50 % LEL
in air, 103 liters |
| 54-9032 |
Propane 50% LEL in
air, 103 liters |
| 54-9033 |
Carbon monoxide (CO),
50 ppm in air, 103 liters |
| 54-9034 |
Hydrogen sulfide
(H2S), 25 ppm in nitrogen, 58 liters |
| 54-9037 |
Sulfur dioxide (SO2),
1 0 ppm in nitrogen, 58 liters |
| 54-9041 |
CO 50 ppm, Methane
50% LEL in air, 103 liters |
| 54-9042 |
CO 50 ppm, Propane
50% LEL in air, 103 liters |
| 54-9043 |
CO 50 ppm, H2S
25 ppm, Methane 50% LEL in air, 58 liters |
| 54-9044 |
CO 50 ppm, H2S
25 ppm, Propane 50% LEL in air, 58 liters |
| 54-9051 |
Ammonia (NH3),
50 ppm in nitrogen, 58 liters. Only for use with 12-025 ammonia
service regulator |
| 54-9052 |
Chlorine (Cl2),
5 ppm in nitrogen, 58 liters |
| 54-9055 |
Nitric oxide (NO),
30 ppm in nitrogen, 58 liters |
| 54-9056 |
Nitrogen dioxide
(NO2), 5 ppm in nitrogen, 58 liters |
| 12-023 |
Regulator, standard
service 1.0 liter per minute flow |
| 12-025 |
Regulator, ammonia
service 1.0 liter per minute flow |
|
PhD
Ultra Accessories
| 35-921 |
PhD Ultra alkaline
battery pack |
| 35-922 |
PhD Ultra NiCad battery
pack |
| 54-05-A0l01 |
Continuous sample
pump |
| 54-05-A0203 |
Slip-in "Datadock"
fast charger |
| 54-05-K0201 |
Datalink software
(includes computer interface cable) |
| 54-05-K0203 |
Datalink Kit (includes
software, interface cable, reference manual, and slip-in "Datadock"
fast charger/computer interface cradle |
| 54-05-A0301 |
Wall-mounting bracket
for PhD Ultra "Datadock" fast charger |
| 54-05-A0205 |
Five instrument slip-in
multi-charger for PhD Ultra |
|
PhD
Ultra Smart Sensors
| 54-04-90 |
Oxygen (02) "smart
sensor" |
| 54-04-80 |
LEL combustible gas
"smart sensor" |
| 54-04-01 |
Carbon monoxide (CO)
"smart sensor" |
| 54-04-02 |
Hydrogen sulfide
(H2S) "smart sensor" |
| 54-04-03 |
Sulfur dioxide (SO2)
"smart sensor" |
| 54-04-04 |
Ammonia (NH3) "smart
sensor" (CALL FACTORY) |
| 54-04-05 |
"CO Plus" dual-purpose
CO/H2S "smart sensor" |
| 54-04-06 |
Nitric oxide (NO)
"smart sensor" |
| 54-04-08 |
Chlorine (CI2) "smart
sensor" |
| 54-04-09 |
Nitrogen dioxide
(NO2) "smart sensor" |
|
WARNING!
Honeywell Analytics - Biosystems equipment should only be used in conjunction
with the manufacturer's instructions.
Failure to follow such instructions could result in serious injury or
fatality.
BACK
|
Return
to Brandt Instruments
For
more information Email
Brandt
Instruments, Inc.
18568 Oak Grove Parkway
Prairieville, LA 70769
1-800-337-6291
Phone: 225-673-6776
Fax: 225-673-6725
Copyright © 2002-2013 Brandt Instruments,
Inc.
Notice:
This publication is intended for general information only and
should not be used as a substitute for reviewing applicable
government regulations, equipment operating instructions, or
legal standards. The information contained in this document
should not be construed as legal advice or opinion nor as a
final authority on legal or regulatory procedures. Brandt Instruments,
Inc. reserves the right to make any changes to information,
features, and specifications of any product without notice.
Brandt Instruments, Inc. has made every effort to ensure the
accuracy of information displayed on this web site, however,
Brandt Instruments, Inc. is not responsible for typographical
errors appearing on this web site. Price and availability information
is subject to change without notice.
|
|
|