Bios DryCal
Primary Gas Flow Calibrators
Industrial Meters - Definer
220, ML-500, ML-800
|
Industrial Precision Air Flow Calibrators
|
|
|
Bios Definer™
220 Primary Gas Flow Calibrator for Standardized Flow Verification
The Definer 220 allows
users to verify gas flow rates, in the field or in the lab, to an accuracy
of ±1% percent of reading, including temperature and pressure compensation.
The Definer 220
is a field-hardy primary flow meter that is NIST traceable. Each Definer
220 is individually calibrated in the Bios ISO 17025-, ANSI Z-540- and
NIST Handbook 150-accredited gas flow calibration laboratory.
Unlike thermal mass flow meters,
the Definer 220's positive displacement technology provides immediate
indication of actual volumetric gas flow rate - accurately and independently
of gas type. Integrated temperature and pressure sensors in the flow
stream allow users to automatically standardize volumetric flow readings
to standard conditions, for traceable verifications of mass flow devices.
The Definer comes
standard with Bios Optimizer™ 110 software.
|
 |
Flow and mass-flow calibrator
with less than 0.5% error, corrects mass flow to Standard Temperature
and Pressure, continuous readout, simple push-button operation, NIST
traceable calibration. Covers 5.0 sccm to 50 L/min with 3 field interchangeable
flow-cells. Uses BIOS's dry-piston technology.
|
 |
The high-speed,
mercury-free, Bios DryCal ML-800 is the most accurate piston prover
with a ±0.15% accuracy of reading, not full-scale. The
Bios DryCal ML-800 is available in three interchangeable flow cells
to cover an ample measuring range, from 5 cc per minute to 50 liters
per minute. Flows are automatically corrected to standardized temperature
and pressure conditions using precision sensors within both the gas
flow stream and the base unit for the highest degree of accuracy. Users
can toggle between volumetric and standardized flow, simultaneously
viewing the flow rate, average, and the number of averaged readings.
Readings may be stored in non-volatile memory, downloaded into a PC
spreadsheet, or printed to the optional battery-powered thermal printer.
|
|
|