Performance
Specifications
| Precision
of Results |
| Wood, Drywall |
±0.05
mg/cm2 |
| Concrete,
Metal |
±0.10
mg/cm2 |
|
Accuracy of Results
|
| Wood, Drywall |
±0.05
mg/cm2 |
| Concrete,
Metal |
±0.10
mg/cm2 |
|
Instrument
Characteristics
| Depth
of Penetration |
No
Interference from objects (lead pipes, etc.) beyond 3/8" |
| Substrate
Effect |
Automatic
Substrate Compensation (Spectrum Analysis) |
| Detection
Technique |
X-Ray
Fluorescence (XRF) |
| Weight |
<3
lb. Total (one piece unit includes 1lb battery pack) |
| Ergonomic |
Grip-handle
for ease of measurements & limited operator fatigue. |
| Measurement
Area |
The
face print is smaller than 1.5" X 1.75". |
| System
battery life |
8
hours of normal usage between charges for each battery |
| Battery
Charge Time |
1.5
hours |
| Operating
Temperature |
20-120°F |
Radiation
Safety Features
| Manual
Shutter Lock |
The
shutter will not open when the manual shutter lock is
activated. |
| Trigger
Lock |
A
key activated trigger lock renders the trigger inoperable. |
| Optical
Sensors |
Ensure
that the bezel of the instrument is flat against the surface
before the shutter can be opened. |
|
|
Radiation
Source
Cobalt-57 (Min. 10
mCi, Max. 12 mCi, 444 MBq)
RMD's LPA-1 analyzer
radioactive source is a 12 mCi, Cobalt 57 isotope. Co-57 gamma
ray energies (122 Kev) are more efficient (at least 10 times)
than those of Cd-109 (22-26 Kev and 88 Kev) source for production
of K-shell lead x-rays, which are recommended by HUD chapter
7 for measurement of lead in paint. Therefore, the measurement
time is shorter using a Co-57 source
Detector
The LPA-1 analyzer utilizes
a "room-temperature" detector, CdTe, for its x-ray analysis
that operates in a wide, 20-120° F, thermal range without
any cooling requirements. The LPA-1 analyzer's ability to operate
without cooling eliminates a cooling down period for stability
at the beginning of a job that would result in extra time added
to inspections.
Operation
The LPA-1 XRF system
has the capability of setting abatement levels from 0.4-2.0
mg/cm2 for automatic determination of the lead condition as
either positive or negative with 95% confidence according to
a PCS approved mode of operation.
Memory
The LPA-1 can data log
up to 4000 data points and is easily down loaded through an
RS-232 serial interface port for report generation with the
WindowsTM based Report Generation Software.
Software
The WindowTM
based interactive graphic software is considered to the most
comprehensive, universally applicable lead paint report generation
system available. The software is included in the LPA-1 instrument
package and upgrades are free for as long as you own the instrument.
Warranty
2-year parts and labor
warranty with a life time warranty on the calibration of the
instrument. RMD Instruments, LLC also offer upgrades of the
software and system upgrades free of charge for as long as you
own your instrument.
All claims of performance
are verified by the Performance Characteristic Sheet, Edition
No.5, December 1, 2006 developed by QuanTech, Inc., under a contract
from the U.S. Department of Housing and Urban Development (HUD). |